Hospital Systems, Inc. Isolated Power Centers for portable X-ray units are designed to provide isolated power to X-ray receptacles located in operating rooms, emergency rooms, and other critical care areas. These Isolated Power Centers include an isolation transformer, a primary and a secondary circuit breaker, a Line Isolation Monitor, an interlocking selector switch station, and power distribution contactors.
Isolation transformers are of low-leakage shielded plate design, available in 7.5, 10, 15, and 25 kVA ratings.
An interlocking selector switch is provided to prevent more than one circuit being activated at a time.
Circuit breakers are 10,000 AIC rated. When typically equipped, the Isolation Power Center includes eight contactors, one main circuit breaker, and one two-pole secondary circuit breaker.
Station switching may be remotely located or may be electrically operated from the receptacle.
Also provided is a high-quality, highly reliable Line Isolation Monitor incorporating dynamic, non-switching circuitry, a micro ammeter, indicator lights, audible device with silencing switch, test switch, and terminals for a remote alarm indicator.
The finish trim is type #304 stainless steel with a #4 brushed finish. The backbox is fabricated from code-gauge steel, finished with durable baked enamel.

Catalog Number:
IPCX-X1-X2/X3-X4-X5-X6
| X1 | Transformer kVA [ 7.5, 10, 15 or 25] |
| X2 | Transformer primary voltage [100, 120, 200, 208, 220, 240, 277 and 480 VAC] |
| X3 | Transformer secondary voltage [100, 120, 200, 208, 220, 240, 277 and 480 VAC] |
| X4 | Number of Secondary Circuit Breakers [1 standard, 8 maximum] |
| X5 | Number of Contactors |
| X6 | 60Hz standard [ 50Hz optional] |
Example: IPCX- 15-240/208-1-8-60 -- designates an Isolation Power Center
with 15 kVA
transformer of 240VAC primary and 208VAC secondary, 60 Hz, with one secondary
breaker and 8
contactors.